• Helium - save it or swap it
    Schematic of Nexis GC-2030 with gas selector.
  • Schematic of Nexis GC-2030 with hydrogen sensor.

Chromatography

Helium - save it or swap it

Jul 15 2022

Solutions for the ongoing helium shortage

Helium is the preferred carrier gas in gas chromatography because of its high inertness and abilities supporting chromatographic separation. The increasing demand for helium in combination with limited availability as well as current events have exacerbated a pre-existing helium shortage. Prices therefore continue to rise, forcing users to look for alternatives.

As alternative carrier gases nitrogen or hydrogen can be used instead of helium. Nitrogen has a sufficient inertness, but lower diffusivity compared to helium or hydrogen. This restricts the use of nitrogen as carrier gas in applications with challenging separation and often the GC run time must be extended to get a sufficient peak resolution.

Hydrogen has a similar diffusivity but roughly two times lower viscosity than helium. It maintains a very good separation efficiency across a wide linear velocity range and therefore hydrogen is the best choice for fast GC applications.

The only disadvantage of hydrogen is the reactivity. It bears an explosion risk if the concentration rises to 4% in the air and forsafe operation strict precautionsmust be taken. TheShimadzu GC-2030 offers a three-step safety solutionfor operation with hydrogen:

  • In case of a big leakage the fast-responding advanced flow controller (AFC) immediately shuts off the carrier-gas supply
  • A preventative, automatic leak-check function helps the operator to check for possible leakages after any kind of column maintenance
  • An independent hydrogen sensor continuously checks the air inside the GC oven and shuts off the carrier gas control once a level of 1% of hydrogen in air is reached

If the GC application does not allow swapping to another carrier gas than helium the GC-2030 offers several functions to reduce the helium consumption:

  • The carrier gas-saver functionality reduces the gas flow during analyses after injection and during brief standby periods
  • 在更长时间的待机期间,气体选择器会切换到氮气。For continuing the measurements, it automatically purges the system for full carrier gas performance with helium
  • Overnight or during weekends and holidays the GC shuts down automatically after the last analysis and restarts in time for the next working day

Depending on standby times and analytical conditions consistent use of these functions the overall gas savings can reach up to 90%.

Key features of the GC-2030 include the free choice of the gas type for carrier gas between helium, hydrogen, nitrogen, and argon. Alternatively, several features and accessories significantly reduce the helium consumption or running cost by switching to nitrogen during standby periods.

Learn more about using hydrogen as carrier gas.